Installation Instructions Gamma 50 LED Light System Corporation w w w . a m i c o .
Contents Symbols 3 Safety Instructions 4 Installation Installation of Gamma 50 Rail Mount Securing the holding device Installing the spring arm Installation of Gamma 50 Ceiling Mount Installing the ceiling fixture Installing the spring arm Operation 11 Disinfection And Cleaning 12 Maintenance 12 Disassembly And Disposal 13 Troubleshooting 13 Technical Data 14 EMC-Inspection Report 2 5 5 6 7 8-9 10 Amico Lighting Solutions 15-17
Symbols2 symbols symbolbeDeutung 2 The warning symbol is used for all instructions that are important for safety. If warnings are ignored, users may be injured or the light or surroundings may be damaged. The warning symbol is used with the following signal words as The warning symbol is used for all instructions that are important for safety. If warnings are Das Warnsym follows: ignored, users may be injured or the light or surroundings may be damaged.
Safety Instructions Intended Use The Gamma 50 LED examination light is designed to provide illumination of the body of a patient for local diagnosis or treatment. The light is not designed for use during surgery. It is designed for continuous operation. It must be possible to stop the treatment or diagnosis without danger to the patient if the light fails during use. User Profiles • Medical staff • All persons with medical training who work in the field for which they were trained.
Installing Gamma 50 LED Rail Mount Included: Gamma 50 LED articulating arm Country-specific power cable Gamma 50 LED gooseneck arm Securing the Holding Device www.amico.
Installing Gamma 50 LED Rail Mount Installing the Spring Arm 6 Amico Lighting Solutions
Installing Gamma 50 LED Ceiling Mount Included: Gamma 50 LED ceiling mount 1x Cable 1x ceiling fixture 6x wall anchors with nuts and washers 1x Plastic ring 1x cover 1x ceiling tube www.amico.
Installing Gamma 50 LED Ceiling Mount Installing the Ceiling Fixture 8 Amico Lighting Solutions
Installing Gamma 50 LED Ceiling Mount www.amico.
Installing Gamma 50 LED Ceiling Mount Installing the Spring Arm 10 Amico Lighting Solutions
Operation Color Blue Color Temperature 4,700 K Ideal for identifying individual vessels for wound follow-up treatment and small procedures 4,100 K Ideal for general examinations 3,500 K Ideal for various types of skin examinations (cold white) White Recommendation (neutral white) Red (warm white) www.amico.
Disinfection And Cleaning Maintenance 12 Amico Lighting Solutions
Disassembly And Disposal Troubleshooting Light does not operate? 1 Switch on light head (keypad) User profiles All 2 Switch on the transformer 3 Connect the mains cable to the transformer casing correct 4 Plug the mains cable to the socket correctly 5 Check mains voltage 6 Contact manufacturer’s service Electricians By manufacturer’s service only www.amico.
Technical Data Electrical: Nominal voltage 100 – 240 V ~ Frequency range 50/60 Hz Power consumption 47 – 56 VA Illumination: Central illuminance (EC) at 0.5 m 50,000 Lux (dimmable) Light field diameter d10 at 0.
EMC-Inspection Report Table 1: Emission Guidance and Manufacturer’s Declaration – Electromagnetic Emissions The Gamma 50 LED is intended for use in the electromagnetic environment specified below. The customer or the user of the Gamma 50 LED should assure that it is used in such an environment. Emission test Compliance RF emissions Electromagnetic environment - guidance Group 1 The Gamma 50 LED uses RF energy only for its internal function.
EMC-Inspection Report Table 3: Immunity (All Devices) Immunity test standard IEC 60601 test level Compliance level Voltage dips, short interruptions and voltage variations on power supply lines <5% UT (0,5 cycle) <5% UT (0,5 cycle) 40% UT (5 cycles) 40% UT (5 cycles) 70% UT (25 cycles) 70% UT (25 cycles) <5% UT for 5s <5% UT for 5s IEC 61000-4-11 Electromagnetic environment guidance Mains power quality should be that of a typical commercial or hospital environment. Note: UT is the a.c.
EMC-Inspection Report Where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in metres (m). Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey, a Should be less than the compliance level in each frequency range.