Owner manual
16 Amico Lighting Solutions
EMC-Inspection Report
Immunity test
standard
IEC 60601 test
level
Compliance level
Electromagnetic environment -
guidance
Voltage dips, short
interruptions and voltage
variations on power
supply lines
IEC 61000-4-11
<5% UT (0,5 cycle)
40% UT (5 cycles)
70% UT (25 cycles)
<5% UT for 5s
<5% UT (0,5 cycle)
40% UT (5 cycles)
70% UT (25 cycles)
<5% UT for 5s
Mains power quality should be that of a typical
commercial or hospital environment.
Note: UT is the a.c. mains voltage prior to application of the test level.
Power frequency
(50/60 Hz)
Magnetic eld
IEC 61000-4-8
3 A/m 100 A/m Power frequency magnetic elds should be at
levels characteristic of a typical location in a typical
commercial or hospital environment.
Guidance and manufacturer’s declaration – electromagnetic immunity
The Gamma 50 LED is intended for use in the electromagnetic environment specied below. The customer or the user of the
Gamma 50 LED should assure that it is used in such an environment.
Electromagnetic environment - guidance
Portable and mobile RF communications equipment should be used no closer to any part of the Gamma 50 LED, including cables,
than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter.
Immunity test standard IEC 60601 test level Compliance level Recommended separation distance C
Conducted RF
IEC 61000-4-6
3 Vrms
150 kHz to 80 MHz
V1 = 10 Vrms
(estimated value)
150 kHz to 80 MHz
d = 0.35 √P 150 kHz to 80 MHz
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz to 800 MHz
E1 = 10 V/m
80 MHz to 800 MHz
d = 0.35 √P 80 MHz to 800 MHz
Radiated RF
IEC 61000-4-3
3 V/m
800 MHz to 2,5 GHz
E2 = 10 V/m
800 MHz to 2,7 GHz
d = 0.7 √P 800 MHz to 2,7 GHz
Table 3: Immunity (All Devices)
Table 4: Immunity (Not Life-Supporting Devices)










