Installation Instructions Skyline Series Light System Corporation
Contents 2 INTRODUCTION 3 GENERAL INFORMATION 3 DRAWINGS AND DIMENSIONS 4 INCLUDED COMPONENTS / PARTS 5 INSTALLING SKYLINE LIGHT Mounting the Hanging Bracket Installing the Fixture Installing the Lamps 6 7 10 WIRING DIAGRAM 11 Amico Lights
Introduction Thank you for purchasing the Skyline overbed light. Incorporating the latest technology the Skyline light is designed to provide high quality illumination while meeting the aesthetic requirements of the patient room. The installation provided are for the Amico standard products. General Information WARNING The installation of the light shall be in accordance with the local building code and the (NEC) National Electrical Code.
Drawings and Dimensions front view top view back view side view 4 Amico Lights
Included Components / Parts Skyline light fixture Hanging Bracket HARDWARE REQUIRED FOR DRYWALL - 3/16” TOGGLE BOLTS FOR WALL STUDS - #10 X ½” PHILLIPS PAN HEAD SELF-DRILLING TEK SCREWS TOOLS SCREWDRIVERS TAPE MEASURE 1/2" DRILL BIT LEVEL WIRE STRIPPER 3/8” SOCKET www.amico.
Installing Skyline Light step 1 Mounting the Hanging Bracket Mounting hardware shall be provided on site by others and shall comply with the local building code. Amico recommends the fixture to be installed at a height of six feet (1.8m) above finished floor. 2" x 4" electrical box should be used for electrical terminations. For feed opening and mounting holes dimensions refer to the drawing of the hanging bracket below.
step 2 Installing the Fixture Remove the top and bottom lens. Remove the power to the electric power supply leads before installation. Read the specified voltage from the fixture label. Be sure the line voltage is the same. To gain access to the wiring compartment remove the top cover. www.amico.
step 2 Installing the Fixture (cont.) Feed wires into the fixture via the back opening. Attach the fixture to the hanging bracket and secure with four #10 Hex Lock Nuts and four washers included with fixtures. Make all necessary connections to the wire terminal blocks according to the wiring diagram provided with the lights, project specific. When complete, replace the cover plate. Replace the top cover. 9 Amico Lights Refer to wire diagrams for specific orientation.
step 3 Installing the Lamps See labels on the reflectors for the type and wattage of the lamps. Install the lamps. Use T5HO 39W lamps for reading and ambient lights. Use T5HO 54W lamps for the direct light. Replace the top and bottom lens. Test light functions, making sure all features are working according to the wiring diagram. Replace lenses. www.amico.
Wiring Diagram 11 Amico Lights
www.amico.com Amico Lights | www.amico.com 85 Fulton Way, Richmond Hill Ontario, L4B 2N4, Canada Toll Free Tel: 1.877.264.2697 Toll Free Fax: 1.866.440.4986 Tel: 905.763.7778 Fax: 905.763.8587 Email: info@amico.com INSTALL INSTR.