CAL:CURR:LEV
This command can only be used in the calibration mode. It sets the power supply to a current calibration
point that is then entered with CAL:CURR[:DATA]. During calibration, three points must be entered and
the low-end point (MIN) must be selected and entered first.
Command Syntax
Parameters
Examples
Query Syntax
Related Commands
CALibrate:CURRent:LEVel <channel > <CRD>
<channel >: 1 – 31, channel number
<CRD>: MIN | MID | MAX
CAL:CURR:LEV 1 MIN; CAL:CURR:LEV 1 MAX
(None)
CAL:CURR[:DATA] CAL:STAT
CAL:CURR:SAV
This commands reloads the current calibration constants to power supply. These constants are not stored in
nonvolatile memory until saved with the *SAV <channel> commands.
Command Syntax
Parameters
Examples
Query Syntax
Related Commands
CALirate:CURRent:SAVe <channel> <NR1>
CURRent:CALibrate:SAVe <channel> <NR1>
<channel>: 1 – 31, channel number
<NR1>: 1 – 12, Calibration mode
1 Voltage output
2 Current output
3 Voltage read back
4 Current read back
CAL:CURR:SAV 1 2; CURR:CAL:SAV 1 4
(None)
*SAV
CAL:OFFS
This command resets or queries the power supply’s calibration offset constant value. These constants are
not stored in nonvolatile memory until saved with the CAL:CURR:SAV or CAL:VOLT:SAV and *SAV
commands.
Command Syntax
Parameters
Example
Query Syntax
Return Parameter
Related Commands
CALibrate:OFFSet <channel> <NR1> <NRf>
SYSTem:OFFSet <channel> <NR1> <NRf>
channel: 1-31, channel number
<NR1>: 1 – 12, calibration mode
1 Voltage output
2 Current output
3 Voltage read back
4 Current read back
<NRf>: offset value
SYST:OFFS 1 1 -0.0665278
CAL:OFFS 1 1 –0.0665278
SYSTem:OFFSet? <channel> <NR1>
CALibrate:OFFSet? <channel> <NR1>
<NRf>: calibration offset constant value
CAL:CURR:SAV; CAL:VOLT:SAV; *SAV