Specifications

CAL:STAT
This command enables and disables the calibration mode. The calibration mode must be enabled before the
power supply will accept any other calibration commands.
Whenever the calibration mode is changed from enabled to disabled, any new calibration constants are lost
unless they have been stored with CAL:SAVE.
Command Syntax
Parameters
*RST Value
Examples
Query Syntax
Returned Parameters
Related Commands
CALibrate:STATe <channel> <bool> [<password>]
<channel>: 1 – 31, channel number
<bool>: 0 | OFF | 1 | ON
<password>: Calibration password.
OFF
CAL:STAT 1 ON; CAL:STAT 1 OFF
CALibrate:STATe? <channel>
0 | 1
CAL:SAV
Note: If the calibration password is set, the password string must match the existing password. Otherwise,
gives an error message.
CAL:VOLT
This command can only be used in the calibration mode. It enters a voltage value that is obtained from an
external meter. Users must first select a calibration level (CAL:VOLT:LEV) for the value being entered.
Two successive values (one for each end of the calibration range) must be selected and entered. The power
supply then computes new voltage calibration constants. These constants are not stored in nonvolatile
memory until saved with the CAL:SAVE command.
Command Syntax
Parameters
Examples
Query Syntax
Related Commands
CALibrate:VOLTage[:DATA] <channel> <NRf>
<channel>: 1 – 31, channel number
<NRf>: External meter value in Volts
CAL:VOLT 1 3.1024; CAL:VOLT 1 5.000
(None)
CAL:SAV; CAL:STAT; CAL:VOLT:LEV
CAL:VOLT:LEV
This command can only be used in the calibration mode. It sets the power supply to a voltage calibration
point that is then entered with CAL:VOLT[:DATA]. During calibration, two points must be entered and
the low-end point (MIN) must be selected and entered first.
Command Syntax
Parameters
Examples
Query Syntax
Related Commands
CALibrate:VOLTage:LEVel <channel> <CRD>
<channel>: 1 – 31, channel number
<CRD>: MINimum |MAXimum
CAL:VOLT:LEV 1 MIN; CAL:VOLT:LEV 1 MAX
(None)
CAL:VOLT[:DATA]; CAL:STAT