Specifications
CAL:VOLT:SAV
This command reloads the voltage calibration constants to power supply. These constants are not stored in
nonvolatile memory until saved with the *SAV <channel> commands.
Command Syntax
Parameters
Examples
Query Syntax
Related Commands
CALirate:VOLTage:SAVe <channel> <NR1>
VOLTage:CALibrate:SAVe <channel> <NR1>
<channel>: 1 – 31, channel number
<NR1>: 1 – 4, Calibration mode
1 Voltage output
2 Current output
3 Voltage read back
4 Current read back
CAL:VOLT:SAV 1 2; VOLT:CAL:SAV 1 4
(None)
*SAV
CHANNEL SUBSYSTEM
The commands in this subsystem only exist in Master – Slave power supply system. They allow users to do
the following:
x Query individual channel’s model information.
x Set or query output shut down time.
CHAN:MOD?
This query returns the model number of a specified channel.
Query Syntax
Parameters
Returned Parameters
Examples
Related Commands
CHANnel:MODel? <channel>
MODel? <channel>
<channel>: 1 – 31, channel number
<CRD> represents model number
CHAN:MOD? 1 MOD? 1
(None)
CHAN:SER
This command sets or queries the power supply’s serial number for a specified channel.
Command Syntax
Parameters
Examples
Query Syntax
Returned parameters
Related Commands
CHANnel:SERial <channel> <CRD>
SERI <channel> <CRD>
<channel>: 1 –31, channel number
<CRD>: 6 characters long
CHAN:SER 1 B12345; SERI 1 B12543
CHAN:SER? <NR1>
SERI? <NR1>
<CRD> represents serial number
(None)
6-21