Specifications

LIST:VOLT:STAT
This command enables or disables the output voltage list operation state.
Command Syntax
Parameters
Examples
Query Syntax
Returned Parameters
Related Commands
LIST:VOLTage:STATe <channel> <bool>
<channel>: 1 – 31, channel number
<bool>: 0 | 1 | OFF | ON
LIST:VOLT:STAT 1 ON; LIST:VOLT:STATE 2 OFF
LIST:VOLTage:STATe? <channel>
0 | 1
LIST:VOLT:STEP
LIST:VOLT:STEP
This command specifies how voltage list sequencing occurs in response to enabling voltage list state
(LIST:VOLT:STAT). If default step (NONE) or AUTO is sent, enabling list operation state causes the
voltage list to sequence through all its points. The time that a list remains at each point is as specified in the
dwell list. As soon as the dwell interval expires, the list moves to the next point. If ONCE is sent, after
enabling list operation state, a single trigger advances a list only one point. After the specified dwell
interval, the list remains at that point until the next trigger occurs.
In either mode, triggers that occur during a dwell interval are ignored.
Command Syntax
Parameters
*RST Value
Examples
Query Syntax
Returned Parameters
Related Commands
LIST:VOLTage:STEP <channel> <CRD>
<channel>: 1 – 31, channel number
<CRD>: NONE | AUTO | ONCE, list step
NONE
LIST:VOLT:STEP 1 ONCE; LIST:VOLT:STEP 2 AUTO
LIST:VOLTage:STEP? <channel>
<CRD>: NONE | AUTO | ONCE, list step
LIST:VOLT; LIST:VOLT:STAT; LIST:VOLT:TIME
LIST:VOLT:TIM
This command sets the dwelling time for the output voltage list of the specified point. Each point specifies
the time, in milliseconds, that the output of the power supply is to remain at the level specified by the
corresponding point in the voltage list if the VOLT:LIST:STEP is set to NONE or AUTO. At the end of
the dwell time, the output of the power supply is set to OFF.
Command Syntax
Parameters
Default Suffix
Examples
Query Syntax
Returned Parameters
Related Commands
LIST:VOLTage:TIMe <channel> <point> <NR1>
<channel>: 1 – 31, channel number
<point>: list point numner
<NR1>: 1 – 99990, dwelling time
ms
LIST:VOLT:TIM 1 10 1000
LIST:VOLTage:TIMe? <channel> <point>
<NR1> dwelling time
LIST:VOLTage:STAT
6-26