Specifications

CC MODE LOW RANGE Calibration (OPTIONAL)
29. CURR <channel> 0.5 - set current less than low range max value (1 Amp)
30. CURR:RANGE <channel> 0 – change to low range current function.
31. CAL:STAT <channel> 2 – enables the calibration mode
32. CAL:CURR:LEV <channel> MIN – power supply sends a low voltage to output.
33. Read voltage value from DMM. Divide value (volts) by the current shunt resistance (ohms) to
obtain calculated current value. Send this value to the power supply.
34. CAL:CURR <channel> value – sends calculated current value to power supply.
35. CAL:CURR:LEV <channel> MID – power supply sends a middle voltage to output.
36. Read voltage value from DMM. Divide value (volts) by the current shunt resistance (ohms) to
obtain calculated current value. Send this value to the power supply.
37. CAL:CURR <channel> value – sends calculated current value to power supply.
38. CAL:CURR:LEV <channel> MAX – power supply sends a high voltage to output.
39. Read voltage value from DMM. Divide value (volts) by the current shunt resistance (ohms) to
obtain calculated current value. Send this value to the power supply.
40. CAL:CURR<channel> value –sends calculated current value to power supply
41. CAL:SAVE <channel> – saves new current calibration data.
42. CAL:STAT<channel. OFF – disables the calibration mode
43. *SAV <channel> – saves all calibration data to internal flash memory
44. Wait about 2-3 minute
Remote Calibration is now complete.