Specifications
CONSTANT CURRENT RANGE SELECTION (OPTIONAL)
The PD features a three range of output operating for Constant Current Mode (CC). The user must select the
PD output range corresponding to the output source value, by pressing the key #6 locates at the front panel
keypad, and then the LCD module will display the operating range L / M / H.
Example:
(H) = 0 – 1 Amp / 30 Volt
(M) = 0 – 15 Amp / 2 Volt
(L) = 0 – 30 Amp / 1 Volt
(L/M/H) Range
STEPS TO VIEW “*IDN?” INFORMATION FROM THE FRONT PANEL
Turn on AC Power or push “Clear” button; the front panel display should contain the following three
fields:
1) AMREL – Manufacturer;
2) PD###### – Model Number;
3) Version number.
STEPS TO SET THE EOS MODE FROM THE FRONT PANEL
1) After front panel display is stable, push “Utility”;
2) Then choose “EOS Mode” by pushing key number 6 “” button;
3) Choose “LF+CR” by pushing key number 2 “” button;
4) Push “Enter” button;
5) Push “Clear” button.
Note: If this is not set to “LF+CR” the output will be ASCII verses a Hex number.
STEP TO VIEW SELF-TEST RESULT
The only way to view the self-test result from front panel is to view the channels from the display. If there is a
“” symbol on top of channel number, it means that self-test for this channel is passed.
PROGRAMMING DC POWER SUPPLY
CV
CC
OVP LCL
OCP RMT
CH RL RNG LIST STS
OUTPUT OFF
01 L