Use Instructions

Table Of Contents
SL5019 Rev. A
Maintenance section for inspection criteria. DO NOT use the rover until it has been tested properly to
ensure functionality. See the To Test the Rover section.
ALWAYS clean the equipment as indicated upon initial receipt and before each use. DO NOT place the
rover within the sterile field. Failure to comply may cause infection and result in patient or healthcare staff
injury.
CAUTION: DO NOT clamp or attach any accessory onto the pole or base of the powered IV pole assembly.
NOTES:
DO NOT use the rover until the docker has been installed and tested properly to ensure functionality. See
the instructions for use supplied with the docker.
ALWAYS close unused manifold ports, and remove or clamp unused tubing to maintain optimal suction
levels.
1. To Connect Power
a. Position the rover on a flat surface and in a convenient location within the operating room.
b. Lock the rover’s two front casters to prevent rover movement.
c. Orient the rover’s touch screen for optimal viewing.
CAUTION: ALWAYS use the correct power cord. Configurations may vary. See the Accessories section for
power cord options.
d. Connect the rover to facility electrical power using the appropriate power cord.
e. Push the power switch to the ON position.
f. Read the WARNING message on the user interface, then tap the OK key to access the CONTROL screen.
g. From the main view, tap the chamber access door keys to Open the canister access doors to allow viewing
of the contents (Figure 16).
NOTE: If the suction setting is in the high suction range, the rover will provide an audible and visual indication of
this condition. See the High Suction Indicator section.
Figure 16 To Connect Power
2. To Prepare For Fluid Suction