Use Instructions

Table Of Contents
SL5019 Rev. A
equipment if damage is apparent or the inspection criteria are not met.
DO NOT disassemble, modify, service, or repair any system component or accessory, unless otherwise
specified. Call AMSINO iReceptal Customer Service.
NOTES:
Only individuals trained and experienced in the maintenance of reusable medical devices should install,
inspect, and test this equipment.
For service, contact your AMSINO sales representative or call AMSINO iReceptal Customer Service.
Outside the US, contact your nearest AMSINO subsidiary.
Maintenance documentation for this equipment is available upon request to AMSINO-authorized service
personnel only.
INTERVAL
INSPECTION CRITERIA ACTION
Before initial use
Check equipment for damage or
missing components and for proper
operation.
If damage is apparent, replace the
equipment. See the Accessories
section.
Make sure the rover and docker
operate as a system properly.
See the following sections: To
Dock the Rover, To Test the
Rover, and To Adjust the Rover
Settings.
Before each use and after
each cleaning
Check equipment for damage or
missing components.
If damage is apparent, replace the
equipment. See the Accessories
section.
Check for cleaning induced damage
or unacceptable deterioration on all
external surfaces of the rover, such as
corrosion, discoloration, pitting, or
cracked materials.
Check the canister and smoke
evacuator filter cover, for cracks or
damage.
Check the four casters with locks and
make sure the locks function properly.
Check the power cord for cuts and the
power cord plug for bent pins.
Check the power cord receptacle for
bent pins or bent contacts.
Six months Check the replacement date on the
smoke evacuator filter label. The
smoke evacuator filter life is 80 hours.
Replace the smoke evacuator
filter every six months or as
indicated on the user interface
display. See the Accessories
section.
NOTE: If any component must be discarded, see the Disposal/Recycle section.