The SUCTION SETTING
are too low.
Adjust the SUCTION SETTING .
The vacuum pump or internal
sensor is damaged.
Remove the rover from use. Contact
AMSINO
iReceptal Customer
Service.
The rover’s suction capability is
lost.
The vacuum pump is damaged. Remove the rover from use. Contact
AMSINO
iReceptal Customer
Service.
A filter error appears on the user
interface display.
The smoke evacuator filter has
exceeded its useful life.
Replace the smoke evacuator filter.
See the Accessories section. See
the instructions for use supplied with
the filter.
The rover is releasing a strong
odor.
The smoke evacuator filter has
exceeded its useful life.
Replace the smoke evacuator filter.
See the Accessories section. See
the instructions for use supplied with
the filter.
The smoke evacuator fails to
operate after activation
(EVACUATE SMOKE key is
Taped).
The smoke evacuator is
damaged.
Remove the rover from use. Contact
AMSINO
iReceptal Customer
Service.
A smoke evacuator error
appears on the user interface
display.
The smoke evacuator filter is not
installed.
Install the smoke evacuator filter.
See the instructions for use supplied
with filter.
The smoke evacuator filter is
installed incorrectly.
Install the smoke evacuator filter
again. See the instructions for use
supplied with filter.
The smoke evacuator is
damaged.
Remove the rover from use. Contact
AMSINO
iReceptal Customer
Service.
IV pole height cannot be
adjusted to its maximum height.
IV pole is supporting too much
weight.
Remove excessive weight from IV
pole.
The IV pole is bent, kinked, or
pinched.
Remove the rover from use. Contact
AMSINO
iReceptal Customer
Service.
Fluid does not drain from the 5-
liter chamber into the 25-liter
chamber after Taping the
EMPTY 4.5L CHAMBER key.
A fluid waste clog exists at the
bottom of the 4.5-liter chamber.
Remove the rover from use.
Close all used and unused manifold
ports.
Turn both SUCTION SETTING dials
to the maximum value. Tap the
SUCTION key. Tap the EMPTY 4L
CHAMBER key.
Contact AMSINO iReceptal