Customer Service if the clog persists.
An error occurs during EMPTY
4.5L
CHAMBER procedure.
More than four EMPTY 4.5L
CHAMBER procedures have been
attempted.
Dock the rover.
The contents of the 4.5-liter
chamber cannot fit into the 25-liter
chamber.
Dock the rover.
Sporadic electrical interference
is experienced.
Electrical noise is present. Turn off all the electrical equipment
not in use in the room.
Relocate the electrical equipment to
maximize the distance between the
equipment. Increase spatial
distance.
Plug equipment into different outlets.
PROBLEM
POTENTIAL CAUSE
CORRECTIVE ACTION
Docking Operation
The rover will not dock or an error
has occurred during docking.
The docker power cord is not
connected or is loosely connected.
Make sure the docker power cord
is connected securely.
The docker power switch is in the
OFF position.
Make sure the power switch is in
the ON position and illuminated.
If power switch is OFF, Tap the
power switch to the ON position.
Wait 60 seconds. Dock the rover.
The docker power switch is in the
ON position, but not illuminated.
Make sure facility power is
provided to the wall receptacle.
If facility power is OFF, apply
facility power to the docker. Wait
60 seconds. Dock the rover.
The docker is not receiving facility
water.
Make sure the water inlet hose is
connected correctly.
Make sure the facility water supply
valve is open.
Dock the rover.
The docker requires a power
reset.
Remove power, then apply power
to the docker. Wait 60 seconds.
Dock the rover.
If the problem persists, the docker
may be damaged.
Contact AMSINO iReceptal
Customer Service.
The docker does not dispense
detergent during the cleaning
cycle.
The bottle of detergent is empty.
Replace the bottle of the detergent.
See the Accessories section.
If the problem persists, the docker
may be damaged.
Contact AMSINO iReceptal
Customer Service.
An error occurs while the rover is
The fluid connectors prevent the
Remove power, apply power to