Use Instructions
Table Of Contents
- Introduction
- Indications For Use
- Contraindications For Use
- For Use With
- Accessories
- System Overview
- User/Patient Safety
- Features
- Physical Buttons
- Graphical user interface
- Definitions
- Instructions
- Inspection and Maintenance
- Cleaning
- Storage and Handling
- Disposal/Recycle
- Troubleshooting
- Error Messages
- Specifications
- Electromagnetic Compatibility
- Appendix
SL5019 Rev. A
54
Pollution Degree 2
Environmental Conditions:
Operation:
Temperature Limitation:
10
℃
~ 40
℃
Humidity Limitation: 30% ~ 75%
Atmospheric Pressure Limitation::70kPa ~ 106kPa
Storage and Transportation (before initial use):
Temperature Limitation:-20℃ ~ 40 ℃
Humidity Limitation: 10% ~ 75%
Atmospheric Pressure Limitation:50kPa ~ 106kPa
Storage and Transportation (after initial use):
Temperature Limitation:10℃ ~ 40 ℃
Humidity Limitation: 10% ~ 75%
Atmospheric Pressure Limitation:50kPa ~ 106kPa
Installation location < 3,000 m above sea level. Usage not allowed in oxygen-
rich or
explosion- hazard environment.
IV Pole Capacity: 12,000 mL or 3000 mL per IV pole hook; for example four three-liter (3000
mL) IV bags
Light Emitting Diode (LED)
Classification (infrared
communication windows):
WARNING: INVISIBLE LED RADIATION
DO NOT VIEW DIRECTLY WITH OPTICAL INSTRUMENTS CLASS
1M LED PRODUCT — Viewing the laser output with certain optical
instruments (for example, eye loupes, magnifiers and microscopes)
within a distance of 100 mm may pose an eye hazard.
Ground Type: Protective Earth (ground); when connected to facility power
Suction
Suction Limit range
37 to 528 mmHg at 1 mmHg increments;measured with all ports closed
5-70 kPa at 1 kPa increments;measured with all ports closed
Suction Tube Connection
Inner diameter:>6.4mm
Suction performance category
High
vacuum/high flow
Vacuum Measurement range 0-80kpa
Vacuum Measurement
Accuracy:
± 5% of full scale
Display resolution
1kPa/1mmHg. Suction level gauge
Fluid collection
Display range
upper chamber 0 …5,000mL
lower chamber 0 … 25,000mL
total volume 0 … 30,000mL
Measurement accuracy 4.5L, ±50ml
22L, ±15
0ml
NOTE: Volume measurements specified do not account for fluid
evaporation or an inclined plane of operation that exceeds the specified
range.