SL5019 Rev. A
55
Smoke evacuation
Maximum Smoke Evacuator Flow
Setting
Standard Hose I.D. Flow Rates
22mm (7/8’’). 707LPM (25CFM)
9.5mm (3/8’’). 130LPM (4.6CFM)
6.4mm (1/4’’). 60LPM (2.1CFM)
NOTE: The specified flow rates were obtained using smoke tubing with a
1.8m length. Actual flow rates may vary depending on the length and
diameter of the smoke tubing used.
Smoke evacuation tube connection
6.4mm, 9.5mm, 22mm
RFID
Operating frequency range
13.56MHz
Channels
Channel separation
Rated maximum peak power
Modulation
FCCID
IC
TABLE 2. Consumables specifications
Single-patient manifold
Shelf life 2 years from manufacturing
Storage Conditions Keep dry, Keep away from sunlight
ULPA filter
Shelf life 2 years from manufacturing
Storage Conditions Keep dry, Keep away from sunlight