Electromagnetic Compatibility
Guidance and manufacturer's declaration - electromagnetic emissions
The iReceptal 3 Rovers, REF iRR301-01 iRR302-01, is intended for use in the electromagnetic environment specified
below. The customer or the user of the iReceptal 3 Rovers, REF iRR301-01 iRR302-01 and, should assure that it is used in
such an environment.
Emissions test
Compliance
Electromagnetic environment - guidance
RF emissions CISPR 11 Group 1 The Rovers, REF iRR301 and REF iRR302, use RF
energy only for its internal function. Therefore, its RF
emissions are very low and are not likely to cause any
interference in nearby electronic equipment.
RF emissions CISPR 11 Class A
The IReceptal 3 Rovers, REF iRR301 and REF
iRR302, are suitable for use in all establishments other
than domestic, and may be used in domestic
establishments and those directly connected to the
public low-voltage power supply network that supplies
buildings used for domestic purposes, provided the
following warning is heeded:
WARNING: This equipment/system is intended for use
by healthcare professionals only. This equipment/
system may cause radio interference or may disrupt
the operation of nearby equipment. It may be
necessary to take mitigation measures, such as
reorienting or relocating the IReceptal 3 Rovers, REF
iRR301 and REF iRR302, or shielding the location.
Harmonic emissions IEC 61000-3-2 Class A
Voltage fluctuations/flicker
emissions
IEC 61000-3-3
Complies
Guidance and manufacturer's declaration - electromagnetic immunity
The IReceptal 3 Rovers, REF iRR301-01 iRR302-01, are intended for use in the electromagnetic environment specified
below. The customer or the user of the IReceptal 3 Rovers, REF iRR301-01 iRR302-01, should assure that it is used in
such an environment.
Immunity test IEC 60601 test level Compliance level Electromagnetic environment - guidance
Electrostatic
dischargeESDIEC
61000-4-2
±6 kv contact
±8 kv air
±6 kv contact
±8 kv air
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%.
Electrical fast
transient/burst IEC
61000-4-4
±2 kv for power supply
Lines
±1kVfor input/output
lines
±2 kv for power supply
Lines
±1kVfor input/output
lines
Mains power quality should be that of a
typical commercial or hospital environment.
Surge IEC 61000-4-
5
±1 kv line(s) to line(s)
±2 kv line(s) to earth
±1 kv line(s) to line(s)
±2 kv line(s) to earth
Mains power quality should be that of a
typical commercial or hospital
environment.
Voltage dips, short
interruptions and
voltage
<5% UT
(>95% dip in UT)
for 0.5 cycle
<5% U
T
(>95% dip in UT)
for 0.5 cycle
Mains power quality should be that of a
typical commercial or hospital
environment. If the user of the IReceptal 3