Figure 42 High Suction Device Warning
3. Install a new disposable manifold into the manifold socket. Make sure the manifold is fully inserted and
locked into place.
4. Attach suction tubing between the hospital wall suction regulator and the auxiliary suction ports on the rover
(Figure 43).
5. Apply and regulate the suction level provided to the rover from the hospital wall suction regulator.
6. Tap the “Auxiliary Suction Port” on the touch screen to have the rover to second control the suction level
(Figure 43).
7. Regulate and monitor suction level.
8. Pressing the Stop Suction button below the touch screen to stop Auxiliary Suction.
Figure 43 To Connect Auxiliary Suction Ports and start Auxiliary suction