Troubleshooting
The FFT plot appears abnormal...
If you see a normal noise floor when you disconnect the signal generator from the
analog input, be sure you are not overdriving the ADC. Reduce input level if
necessary.
In VisualAnalog, Click on the Settings button in the “Input Formatter” block. Check
that “Number Format” is set to the correct encoding (offset binary by default). Repeat
for the other channel.
If DC Spurs appear in the “FFT” plot from VisualAnalog? See Figure 2.
Figure 2: DC Spurs on FFT
Add a ferrite core to ALL switching power supplies used in the evaluation platform,
like Digi-Key Part number: 240-2287-ND (M/N: LFB360230-300). See Figure 3.
DC Spurs