Use Instructions

Table Of Contents
15
Version C
Clinician Instructions for Use Device and Mobile Application
Respiration Rate/Tidal Volume (RR/TV)
This is a metric that computes the ratio of respiration rate and relative tidal volume. An upward trend can indicate an
increase in respiration rate and/or a decrease in tidal volume, suggesting a rapid shallow breathing pattern.
QRS Width
QRS width, measured in milliseconds (ms), represents ventricular depolarization. It is
calculated as the difference between QRS onset and offset, which is determined
based on a metric representing the QRS-like morphology in the single-lead ECG
waveform. The reported value is the mean value from the lowest-angle posture.
QT Interval and QTc
The QT interval is the time from the beginning of the QRS complex to the end of the T
wave resulting from ventricular repolarization. QTc is the corrected QT interval,
accounting for the effects of heart rate. It is calculated using the equation
QTc=QT/

. QT interval and QTc are both measured in milliseconds (ms) and are
calculated from the ECG waveform.
Potential cardiac abnormalities
The CPM System is intended to detect and flag ECG tracings demonstrating an irregularly irregular rhythm such as
that seen in an atrial fibrillation.
Other types of arrhythmias not exhibiting this characteristic are
NOT
intended to be detected and will not be flagged
by the CPM System. An example of an abnormal rhythm flag can be found below.
NOT DETECTED
Normal Sinus
Rhythm
DETECTED
“Abnormal Rhythm”
WARNING
This device is
NOT
intended to be used as an apnea monitor. Do not rely on the respiration
monitoring function of the device for detection of the cessation of breathing.