Use Instructions
Table Of Contents
- GLOSSARY & ACRONYMS
- PART 1. GENERAL SYSTEM INFORMATION
- 2. INDICATIONS AND CONTRAINDICATIONS FOR USE
- 3. WARNINGS AND PRECAUTIONS
- CAUTION
- 6. BASELINE MEASUREMENT PREPARATION AND DEVICE SETUP
- 8. THE ALIGNMENT TOOL
- PART 3. AT-HOME CPM SYSTEM USE
- PART 4. LIGHTS, SOUNDS, AND TROUBLESHOOTING
- 12. CPM DEVICE LIGHTS AND SOUNDS
- 13. BASE STATION INDICATIONS
- 14. BUTTON PRESSES
- 15. GENERAL TROUBLESHOOTING / FAQS
- 15.1 Physician Q&As
- Q: No data from my patient is appearing on the server, but after getting in touch with the patient, they say that they have been taking measurements consistently with no issue and their device is connected to a plugged- in Base Station.
- Q: No size of the CPM Device fits my patient.
- Q: I initiated a patient mode measurement while trying to put the device into Bluetooth pairing mode and now I can’t pair the CPM Device to the Mobile App.
- Q: How do patients get more adhesives when they run out?
- Q: Can the CPM System be used near other electronic or medical equipment such as telephones, televisions, computers, etc.?
- Q: Is data secure when transmitting from the CPM Device to the Mobile App, the Device to the Base Station, or the Base Station to the Cloud platform?
- 15.2 Patient Q&As
- Q: The CPM Device is not lighting up when plugged into the Base Station to charge.
- Q: Nothing happens when I am trying to start a measurement and/or the button is pressed on the CPM Device.
- Q: The device visually does not stick well to my body or is obviously peeling up.
- Q: I lost the clear protective liners for covering the adhesive.
- Q: I pressed the button accidentally/too early and a measurement started before the device was on my body. How do I get the device to stop beeping?
- Q: The CPM Device continues to beep, indicating there are errors, during a measurement.
- Q: Can I shower with the CPM Device or get it wet?
- Q: Can I travel with the CPM System?
- Q: A piece or part broke off the device, base station, or alignment tool.
- Q: What do I do if I think I am having a medical emergency or decompensation event?
- 15.1 Physician Q&As
- PART 5. CLEANING AND MAINTENANCE
- PART 6. SYSTEM SPECIFICATIONS
- PART 7. SYMBOLS GLOSSARY
29
Version C
Clinician Instructions for Use – Device and Mobile Application
6)
When Position 1 is complete, review the data and proceed using the “Looks Good
Continue” button.
7)
Move the patient into your desired Second Position.
IMPORTANT
The default for Position Two is at least 50 degrees less than Position
One. You can override this default by selecting “Skip and Continue”
when the tilt error pop-up box appears.
8) Select the “Continue Reading” button and repeat the steps above for the second position.
When you are reviewing your reading, ensure that “Reading is a Baseline” is checked at
the top of the screen.
Note: If you would like to save a reading taken with the mobile application as a non-
baseline, normal reading, uncheck the “Reading is a Baseline” button before saving.
IMPORTANT
Do not plug the device in to the Base Station until you complete the
ENTIRE Baseline workflow
. Otherwise the CPM Device will disconnect
from the application and you will not be able to enter thresholds and/or
patient details from the app.
9) Review and Confirm the thresholds. These thresholds should match the clinical
needs of the patient and will only apply to this specific patient. They will also appear
and be editable on the Web Application. Default thresholds should already be set by your
organization’s policy. Then select “Confirm Settings and Continue”.