Use Instructions

Table Of Contents
29
Version C
Clinician Instructions for Use Device and Mobile Application
6)
When Position 1 is complete, review the data and proceed using the “Looks Good
Continue” button.
7)
Move the patient into your desired Second Position.
IMPORTANT
The default for Position Two is at least 50 degrees less than Position
One. You can override this default by selecting “Skip and Continue”
when the tilt error pop-up box appears.
8) Select the “Continue Reading” button and repeat the steps above for the second position.
When you are reviewing your reading, ensure that “Reading is a Baseline” is checked at
the top of the screen.
Note: If you would like to save a reading taken with the mobile application as a non-
baseline, normal reading, uncheck the “Reading is a Baseline” button before saving.
IMPORTANT
Do not plug the device in to the Base Station until you complete the
ENTIRE Baseline workflow
. Otherwise the CPM Device will disconnect
from the application and you will not be able to enter thresholds and/or
patient details from the app.
9) Review and Confirm the thresholds. These thresholds should match the clinical
needs of the patient and will only apply to this specific patient. They will also appear
and be editable on the Web Application. Default thresholds should already be set by your
organization’s policy. Then select “Confirm Settings and Continue”.