Use Instructions
Table Of Contents
- GLOSSARY & ACRONYMS
- PART 1. GENERAL SYSTEM INFORMATION
- 2. INDICATIONS AND CONTRAINDICATIONS FOR USE
- 3. WARNINGS AND PRECAUTIONS
- CAUTION
- 6. BASELINE MEASUREMENT PREPARATION AND DEVICE SETUP
- 8. THE ALIGNMENT TOOL
- PART 3. AT-HOME CPM SYSTEM USE
- PART 4. LIGHTS, SOUNDS, AND TROUBLESHOOTING
- 12. CPM DEVICE LIGHTS AND SOUNDS
- 13. BASE STATION INDICATIONS
- 14. BUTTON PRESSES
- 15. GENERAL TROUBLESHOOTING / FAQS
- 15.1 Physician Q&As
- Q: No data from my patient is appearing on the server, but after getting in touch with the patient, they say that they have been taking measurements consistently with no issue and their device is connected to a plugged- in Base Station.
- Q: No size of the CPM Device fits my patient.
- Q: I initiated a patient mode measurement while trying to put the device into Bluetooth pairing mode and now I can’t pair the CPM Device to the Mobile App.
- Q: How do patients get more adhesives when they run out?
- Q: Can the CPM System be used near other electronic or medical equipment such as telephones, televisions, computers, etc.?
- Q: Is data secure when transmitting from the CPM Device to the Mobile App, the Device to the Base Station, or the Base Station to the Cloud platform?
- 15.2 Patient Q&As
- Q: The CPM Device is not lighting up when plugged into the Base Station to charge.
- Q: Nothing happens when I am trying to start a measurement and/or the button is pressed on the CPM Device.
- Q: The device visually does not stick well to my body or is obviously peeling up.
- Q: I lost the clear protective liners for covering the adhesive.
- Q: I pressed the button accidentally/too early and a measurement started before the device was on my body. How do I get the device to stop beeping?
- Q: The CPM Device continues to beep, indicating there are errors, during a measurement.
- Q: Can I shower with the CPM Device or get it wet?
- Q: Can I travel with the CPM System?
- Q: A piece or part broke off the device, base station, or alignment tool.
- Q: What do I do if I think I am having a medical emergency or decompensation event?
- 15.1 Physician Q&As
- PART 5. CLEANING AND MAINTENANCE
- PART 6. SYSTEM SPECIFICATIONS
- PART 7. SYMBOLS GLOSSARY
32
Version C
Clinician Instructions for Use – Device and Mobile Application
C
3)
Once the housing attachment mechanism is in place, push the q
release slider down to lock the component securely in place on t
CPM Device.
4)
Gently press the protruding bulb of the flexible strap into the pati
clavicular notch, fitting it comfortably into the bottom of the notch.
uick
he
ent’s
CAUTION
Bruising can occur when the alignment tool is applied
with excessive pressure at the base of the clavicular
notch. Take special precaution for those patients who
are on blood thinners or steroids.
5)
Visualize, count, or mark where the strap meets the top of the
sn
on mechanism.
6)
Using scissors, cut the alignment strap at the notch below the
po
you just identified. Discard the piece below the cut. You may wan
cut a longer than necessary length at first and then trim.
ap
-
int
t to
7)
Slide the alignment strap into the slot at the top of the
snap-on mechanism.
8)
Once satisfied with the fit of the strap, move the blue
slider up to release the snap-on mechanism from the
electronics housing.
9)
Remove the silver pin from the locking snap and press the lockin
snap down firmly with the alignment strap inserted into the slot to
permanently secure the alignment guide inside the housing
attachment mechanism.
10)
The alignment tool should fit as shown below.
g