Clinician Instructions for Use Device and Mobile Application
12.3 Notifications and alerts during a measurement
Description
Active reading is taking place
CPM Device is currently
recording data. A tone will play
every 5 seconds.
Move to the next position
Error resolved /
measurement position
beginning / measurement
complete
Sounds
Single note every 5 seconds
3-note tone
2 tones
Actions
Remain still and quiet for the
duration of the measurement
and do not touch the sensors.
Move to the next position
specified by the Clinical Care
Team and press the blue button
on the electronics housing to
start the second half of the
measurement.
Description
Tilt Error Lean forward
You are leaning too far back
and need to lean forward.
Tilt Error Lean backward
You are leaning too far forward
and need to lay back further
Electrode Contact Error
The electrodes are not
contacting the skin well
Sounds
3 ascending notes
3 descending notes
4 flat notes
Actions
Lean forward until the LED
flashes green and the error
goes away; otherwise it will
resolve after 20 seconds.
Lean backward until the LED
flashes green and the error
goes away; otherwise it will
resolve after 20 seconds.
Press down on the parts of the
device with the electrodes until
LED flashes green and the error
resolves. You may also have to
replace the adhesives, dampen
skin, or remove and replace the
device.