Use Instructions
Table Of Contents
- GLOSSARY & ACRONYMS
- PART 1. GENERAL SYSTEM INFORMATION
- 2. INDICATIONS AND CONTRAINDICATIONS FOR USE
- 3. WARNINGS AND PRECAUTIONS
- CAUTION
- 6. BASELINE MEASUREMENT PREPARATION AND DEVICE SETUP
- 8. THE ALIGNMENT TOOL
- PART 3. AT-HOME CPM SYSTEM USE
- PART 4. LIGHTS, SOUNDS, AND TROUBLESHOOTING
- 12. CPM DEVICE LIGHTS AND SOUNDS
- 13. BASE STATION INDICATIONS
- 14. BUTTON PRESSES
- 15. GENERAL TROUBLESHOOTING / FAQS
- 15.1 Physician Q&As
- Q: No data from my patient is appearing on the server, but after getting in touch with the patient, they say that they have been taking measurements consistently with no issue and their device is connected to a plugged- in Base Station.
- Q: No size of the CPM Device fits my patient.
- Q: I initiated a patient mode measurement while trying to put the device into Bluetooth pairing mode and now I can’t pair the CPM Device to the Mobile App.
- Q: How do patients get more adhesives when they run out?
- Q: Can the CPM System be used near other electronic or medical equipment such as telephones, televisions, computers, etc.?
- Q: Is data secure when transmitting from the CPM Device to the Mobile App, the Device to the Base Station, or the Base Station to the Cloud platform?
- 15.2 Patient Q&As
- Q: The CPM Device is not lighting up when plugged into the Base Station to charge.
- Q: Nothing happens when I am trying to start a measurement and/or the button is pressed on the CPM Device.
- Q: The device visually does not stick well to my body or is obviously peeling up.
- Q: I lost the clear protective liners for covering the adhesive.
- Q: I pressed the button accidentally/too early and a measurement started before the device was on my body. How do I get the device to stop beeping?
- Q: The CPM Device continues to beep, indicating there are errors, during a measurement.
- Q: Can I shower with the CPM Device or get it wet?
- Q: Can I travel with the CPM System?
- Q: A piece or part broke off the device, base station, or alignment tool.
- Q: What do I do if I think I am having a medical emergency or decompensation event?
- 15.1 Physician Q&As
- PART 5. CLEANING AND MAINTENANCE
- PART 6. SYSTEM SPECIFICATIONS
- PART 7. SYMBOLS GLOSSARY
50
Version C
Clinician Instructions for Use – Device and Mobile Application
C
PART 6. SYSTEM SPECIFICATIONS
19.
TECHNICAL SPECIFICATIONS
MEASUREMENT PARAMETERS
Impedance
Range:
50 – 250 Ohms
Accuracy:
0.5 Ohms
RMS Current Injection:
9 uA
Respiration Rate Derived Range:
6 – 40 breaths per minute (+/- 1 breath per minute)
ECG
Sampling Rate:
1 kHz
Digital Resolution:
16 bits
Input Dynamic Range:
+/- 5 mV
Input Offset Dynamic Range:
+/- 300 mV
Heart Rate Derived Range:
30 – 200 beats per minute (+/- 5 beats per minute)
Heart Sounds
Sampling Rate:
5 kHz
Diginal Resolution:
16 bits
Frequency Range:
10 Hz – 1 kHz
Temperature
Range:
34 – 43 °C
Precision:
0.3 deg °C
Tilt
Range:
0° – 360°
Precision:
5°