Installation Instructions
ADI Confidential 
Rev. Sp0 | Page 17 of 20 
BATTERY LIFE AND MEMORY FOOTPRINT 
BATTERY LIFE 
The battery life is determined by the type and number of 
sensors enabled, the sampling rate configured and is also 
dependent on other configuration details such as the LED 
current in the photoplethysmography measurements. 
Table 10. System Battery Life for Several Use Cases 
Use Case 
Memory 
Live 
BLE 
Battery  
Life 
(Hours) 
Shipment 
Mode 
No 
No 
30000+ 
Hibernate 
No 
No 
1000+ 
PPG – High 
Performance 
Yes 
Yes 
14  
ECG – High 
Performance 
Yes 
Yes 
40 
PPG + 
Impedance 
Yes 
Yes 
50 
PPG + ECG 
Yes 
Yes 
40 
PPG – All LEDs 
Yes 
Yes 
20 
The VSM watch is powered by a rechargeable 200 mAh 
battery. 
MEMORY FOOTPRINT 
The watch hosts a 512 MB memory. 
The actual capacity of the memory is determined by the type 
and number of sensors enabled and the sampling rate 
configured. 
Table 11. Memory Capacity (in Hours) for Different Use 
Cases 
Vital Signs 
Memory Capacity 
(Hours) 
PPG – High Performance 
18 
ECG – High Performance 
22 
PPG + Impedance 
63 
PPG + ECG 
43 
PPG – All LEDs 
23 
Note : All the default usecases are designed to maximize 
the performance. By modifing various parameters of the 
AFEs one can achieve longer battery life and memory usage. 
For instance in ADPD4100 the default gain in AGC algorithm 
is set to 25K so as to get best PPG performance with various 
skin tones. Setting the gain to 100k/200k and lowering 
current value will result in significant improvement in 
battery life. 