User guide

4 Start-Up
4.1 Contents of Shipping Container:
4.1.1 AII-3000 A, AII-3000 M:
ENCL-1061 V-mount retainer (attached)
ENCL-1066 Tripod wire stand (attached)
AII-11-60 Oxygen Sensor
BATT-1008 Battery, AA 1.5V Alkaline (Qty 2)
CABL-1006 Cable, Coiled Phone Jack
P-1087 Instructions for Use
4.1.1 AII-3000 AHC, AII-3000 MHC:
ENCL-1061 V-mount retainer (attached)
ENCL-1066 Tripod wire stand (attached)
AII-11-60-HC Oxygen Sensor (installed inside analyzer)
BATT-1008 Battery, AA 1.5V Alkaline (Qty 2)
TUBE-1019 Tubing, 7/32OD Tubing 3 ft.
P-1087 Instructions for Use
Inspect the box and contents for shipping damage. If the device or
components appear damaged, do not attempt to operate the device -
contact the manufacturer immediately, refer to section 9.
6
4.2 Install Batteries
All devices are powered by two 1.5V AA alkaline batteries which must be in-
stalled before the device can be operated.
The battery compartment is located at the rear of all devices. Initially
this procedure can be somewhat difficult. Care should be taken not to
damage the case when removing the battery compartment cover.
4.2.1 Procedure:
1. Remove the device and the (2) AA 1.5V Alkaline batteries from the foam
shipping container.
2. Turn the device over so the shortest raised line on the battery compart-
ment cover is pointing away from you.
3. Lift the tripod wire stand up and away from the case.
4. Grasp the case with both hands, use your thumbs press down firmly on
the raised lines and push the battery compartment cover away from you.
5. Locate the positive (+) and negative (-) terminals on the battery.
6. Assure the battery contacts are clean.
7. Align one battery’s positive (+) terminal with the corresponding (+) bat-
tery symbol molded into the case.
8. Insert the battery into the compartment.
9. Repeat with the remaining battery.
10. Replace the battery compartment cover, make sure it snaps into position
and is secured flush against the case. Replace the wire stand as required.
Replace the batteries within twenty-four (24) hours of the battery sym-
bol appearing on LCD display because batteries decline at different
rates. Calibrate the device after replacing the batteries.
7