Advanced Instruments Inc.
2 Quality Control Certification
Date: Customer: Order No.: Pass
Model
GPR-3100 High Purity % O2 Analyzer
Sensor
GPR-11-120-OP
Accessories
Owner’s Manual
CABL-1008 Power Cord Black 3 Conductor UL/CSA
HRWR-1021 Wrench 5/16 Combination
Configuration
Ranges: 0-100%; suppressed 50-100%, 80-100%, 90-100%
A-1146-E-50 PCB Assembly Micro-processor / Display
A-1147-E-50 PCB Assembly Power Supply / Interconnection
Software Rev: _______
Wetted parts: Stainless steel
Power: Specify: ( x ) 100/120 VAC, ( ) 220/240 VAC
Enclosure: ( ) Panel mount 10.8”W x 7.5”H
( ) Rack mount panel 19”W x 12”H
( x ) Wall mount 12”W x 12”H x 8”D
Test - Electronics
Factory default zero (without sensor)
Factory default span @ 20 uA
Alarm relays activate/deactivate to changes in O
2
concentration
Sensor failure alarm
Power failure alarm
Analog signal output 0-1V and 4-20mA
Range ID voltage output
Test - Gas Phase
Baseline drift < ± 1% FS 90-100% range over 24 hour period
Noise level < ± 1.0% FS
Span adjustment within 10-50% FS
Temperature controller (CNTL-_______) set at 85°F
Final
Overall inspection for physical defects
Notes
1 of 1 analyzer due ASAP
GPR-3100 Rev 7/2004 2 1 of 1