BREATHALYSER
Congratulations on buying your new Wingmate Pro breathalyser. We hope you enjoy owning your Wingmate Pro breathalyser. If you have any questions regarding the unit or want to speak with us further, please visit andatech.com.au or call 1300 800 200 (between 9am and 5pm Monday to Friday).
Display Battery low. Replace with new batteries. Device is ready to take a test. Start blowing. Test result is displayed as a 3 decimal place %BAC reading Displays test result Result is above 0.050% BAC Dual colour LCD backlight. See Warnings Display section in the manual. Operating Instructions 1. Insert a mouthpiece into the mouthpiece slot 2. Press the power button. “ON” will display and the breathalyser will start warming up, which will take about 15 ~ 35 seconds. 3.
Warnings Display The breathalyser’s LCD Backlight will turn red when any of these errors occur. If the result is higher than 0.050% BAC, the breathalyser will display “WARN” and sound an audible alert. If the result is above 0.400% BAC, “Hot” signal is displayed along with alarm sound. If you have tested over 500 times, CAL will show up on the display screen for 3 seconds before warming up for a test. Please send the breathalyser back for calibration. The battery voltage is too low to perform a test.
unit Pro The Wingmate Pro uses a fuel cell sensor to measure breath alcohol levels. Depending upon the amount of alcohol registered by the sensor after a period of time, it will require recalibration. Andatech recommends the Wingmate Pro is calibrated every 6 months or after 500 uses, whichever comes first. To maintain the unit to Australian Standards, calibration every one (1) month is required.
Version 1.0. Updated December 2015.
Product Registration Wingmate Pro