User Manual

Page 16 of 27
Hold the meter and move the absorbent hole of the test strip to catch
the drop. Once the confirmation window fills completely, the meter will
start counting down. Please leave the test strip blood samples when the
countdown begins
Note: To avoid contaminating the entire vial of Control Solution, do not directly
apply Control Solution onto a strip.
Step 5: Read and Compare the Results
After counting down to 1 the Control Solution test result
will appear on the meter display.
The result of the Control Solution test should be within the
range printed on the test strip vial label. If the test result falls outside the
specified range, repeat the test, carefully following the steps above.
I Performing an Online Control Solution Test (with iOS device)
Step 1: Launch the App
Step 2: Tap the “CTL” button to switch to Control (“CTL”) mode
Step 3: Follow Steps 1-4 of “Performing an Offline Control Solution Test”
Step 4: the Control Solution test result will appear on the iOS device.
Out-of-Range Results
Results falling outside the specified range may be caused by:
Error in performing the test
Expired or contaminated Control Solution
Expired or contaminated test strip
iHealth Wireless Smart Gluco-Monitoring System (BG5) malfunction
If you continue to get Control Solution test results that fall outside of the range
printed on the vial, the meter may not be working properly. Discontinue use
and call iHealth Labs Customer Service at +1855-816-7705 for assistance.
To purchase additional Control Solution, call iHealth Labs Customer Service
at +1855-816-7705.
NOTE:
Do not use expired Control Solution.
The control solution range printed on the test strip vial is for Control Solution
use only. It is not a recommended range for your blood glucose level.