User Manual

Page 20 of 27
1 After the test, the user should clean and wash hands, and then put on the disposable glove.
2 Use the Cavi WipesTM (Metrex® RESERCH CORPORATION,EPA Reg. NO. 467818, EPA Est. No.
56952WI001) to clean and disinfect the meters housing and lens,.Take one wipe to clean the
meter, including its front and back for 10 seconds for cleaning until there is no soil on the surface.
Then disinfect the meter with another wipe and keep the surface wet for 2 min.
3 With the same method use the Cavi WipesTM (Metre RESERCH CORPORATION,EPA Reg. NO.
467818, EPA Est. No. 56952WI001) to clean and disinfect the lancing device..
4 Take off the glove and the protective cover
Note:
1. Each disinfection step requires a precleaning step.
2. The users should wash hands thoroughly with soap and water after handling the meter, lancing
device, or test strips.
3. The meter can only be cleaned and disinfected with the disinfecting towelette on the surface of the
device, users can not insert the disinfecting towelette into the test strip port and the metal contector,
and else the performance of the meter may be influenced.
4. Please refer to the following linkage for detail information: “FDA Public Health Notification: Use of
Fingerstick Devices on More than One Person Poses
Risk for Transmitting Bloodborne Pathogens: Initial Communication” (2010)
http://www.fda.gov/MedicalDevices/Safety/AlertsandNotices/ucm224025.htm
“CDC Clinical Reminder: Use of Fingerstick Devices on More than One Person Poses Risk for
Transmitting Bloodborne Pathogens” (2010)
http://www.cdc.gov/injectionsafety/FingerstickDevicesBGM.html
5. If you have any questions you can call the number+18558167705
6. Table: A list of the signs of potential deterioration
(Physical and in performance)
What do you see?
What to do?
The device does not workfor
example the iOS device can’t begin
testing when meter is connect with
iOS device or when test strip is
Stop using and contact local customer services or place of
purchase for assistance, or toll free service hotline:
+18558167705