Test results are
inconsistent or Control
Solution test results are
not within the specified
range.
1. Not enough sample
in the Test Strip.
2. Test strip or Control
Solution has expired.
3. Test strip has been
damaged due to heat
or humidity so that
the sample cannot be
applied, or the speed
of application is too
slow.
4. System is not
performing due to the
environment being
above or below room
temperature.
1. Re-test with a new
Test Strip and make
sure that enough
sample has been
applied.
2. Re-test with a new
Test Strip or new
Control Solution
3. Perform a Control
Solution test
using a new Test
Strip. If the results
are still out of
range, replace
with new vial of
Test Strips.
4. Bring the system
to a
room-temperature
environment and
wait approximately
30 minutes before
performing a new
test.
The meter countdown
did not start.
Test strip has not been
inserted correctly.
Use a new Test
Strip and redo the
test.