User Manual

Note: These specifications are subject to change without notice.
1. Read all of the information in the Owner’s Manual and other provided instructions before
operating the unit.
2. Consult your physician for any of the following situations:
a) The application of the cuff over a wound or inflamedarea.
b) The application of the cuff on any limb with intravas cular access or therapy, or an arterio-
venous (A-V) shunt.
c) The application of the cuff on the arm on the side of a mastectomy.
d) Simultaneous use with other medical monitoring equipment on the same limb.
e) The blood circulation of the user needs to be checked.
GENERAL SAFETY AND PRECAUTIONS
9 10
3. This Wireless Blood Pressure Monitor is designed for adults and should never be used on
infants, young children, pregnant or pre-eclamptic patients. Consult your physician before
use on children.
4. Do not use this product in a moving vehicle as this may result in inaccurate measurements.
5. Blood pressure measurements determined by this product are equivalent to those obtained
by professional healthcare practitioners using the cuff/stethoscope auscultation method
within the limits prescribed by the American National Standard, Electronic or Automated
Sphygmomanometer.
6. Information regarding potential electromagnetic or other interference between the blood
pressure monitor and other devices together with advice regarding avoidance of such
interference, please see ELECTROMAGNETIC COMPATIBILITY INFORMATION. It is
suggested that the blood pressure monitor be kept 10 meters away from other wireless
devices, such as WLAN unit, cell phone, microwave oven, etc.
7. It is recommended that the iOS device be set in Airplane mode during measurement to
avoid strong magnetism interference. If a call comes in during the measurement, the
measurement process will be terminated automatically.
8. If Irregular Heartbeat (IHB) is detected during the measurement procedure, the IHB symbol
will be displayed. Under this condition, the Wireless Blood Pressure Monitor can keep
functioning, but the results may be inaccurate. Please consult your physician for accurate
assessment.
There are 2 conditions under which the signal of IHB will be displayed:
a) The coefficient of variation (CV) of pulse period >25%.
b) The difference of adjacent pulse period ≥0.14s and the number of such pulse takes more
than 53 percent of the total number of pulses.
9. Please do not use any other cuff other than that supplied by the manufacturer as this may
!
18. Battery life: more than 80 measurements on a full charge
19. The blood pressure measurement system includes accessories: pump, valve, cuff, and sensor
11. Accuracy:
Pressure: ±3 mmHg
Pulse rate: ±5%
12. Wireless communication: Bluetooth V3.0 + EDR Class 2 SPP
Frequency Band: 2.402-2.480 GHz
13. Environmental temperature for operation: 5°C~40°C(41°F~104°F)
14. Environmental humidity for operation: ≤90%RH
15. Environmental temperature for storage and transport: -20°C~55°C(-4°F~131°F)
16. Environmental humidity for storage and transport: ≤95%RH
17. Environmental pressure: 80kPa-105kPa