User Manual

13 14
1. Do not drop this monitor or subject it to strong impact.
2. Avoid high temperature and direct sunlight. Do not immerse the monitor in water as this will
result in damage to the monitor.
3. If this monitor is stored near freezing temperatures, allow it to acclimate to room temperature
before use.
4. Do not attempt to disassemble this monitor.
5. If the monitor is not used for a long time, please sure to fully charge it every month.
6. It is recommended that product performance be checked every 2 years or after each repair.
Please contact the Customer Service.
CARE AND MAINTENANCE
TROUBLESHOOTING
!
!
!
PROBLEM POSSIBLE CAUSE SOLUTION
Battery is less than 20% Charge the batteryLow Battery
Blood pressure is outside of measurement range Retest, make sure your blood pressure is within
measurement range
Arm or monitor was moved during test Retest, make sure not to move your arm or the monitor
The cuff does not inflate properly or pressure
falls quickly during test
Review the cuff application instructions and retest
Irregular heartbeat (arrhythmia) It is inappropriate for people with serious arrhythmia to
use this monitor. Check with physician
The cuff was not properly
applied
Review the cuff application instructions and retest
Display reads
“ERROR”
The cuff position was not correct or it was not
properly tightened
Review the cuff application instructions and retest
Incorrect operation or strong electromagnetic
interference
Press the START/STOP button about 10 seconds to reset
the device, relaunch app, and reconnect the iOS device
to the monitor
No response
Body posture was not correct during testing Review body posture instructions and retest
Speaking, moving arm or body, being angry,
excited or nervous during test
Retest when calm; avoid speaking or movement
during the test
Bluetooth connection unsuccessful, monitor is
abnormal, or strong electromagnetic
interference is present
Reset iOS device. Reset monitor by pressing the
START/STOP button about 10s. Make sure the monitor
and iOS device are away from other electrical
equipment. Please see GENERAL SAFETY AND
PRECAUTIONS.
Display reads
an abnormal
result
Bluetooth
connection
unstable
7. No monitor component needs to be maintained by the user. The circuit diagrams, component
part lists, descriptions, calibration instructions, or other information which will assist the user’s
appropriately qualified technical personnel to repair those parts of the equipment which are
designated for repair can be supplied.
8. Clean the monitor with a dry, soft cloth or a moistened and well wrung soft cloth using water,
diluted disinfectant alcohol, or diluted detergent.
9. The monitor can maintain the safety and performance characteristics for a minimum of 10,000
measurements or three years of usage.
10. The battery can maintain the performance characteristics for a minimum of 300 charge cycles.
Battery replacement should only be performed by a qualified iHealth technician. To will void
your warranty and possibly damage your unit.
11. Cuff replacement should only be performed by a qualified iHealth technician. To do otherwise
will possibly damage your unit.