User`s guide

FUNCTIONS
SDK SECTION 11
Page 196
SetBackground
unsigned int WINAPI SetBackground(at_32* arr, unsigned long size)
Description THIS FUNCTION IS RESERVED.
SetBaselineClamp
unsigned int WINAPI SetBaselineClamp(int state)
Description
This function turns on and off the baseline clamp functionality. With this feature enabled
the baseline level of each scan in a kinetic series will be more consistent across the
sequence.
Parameters
int state: Enables/Disables Baseline clamp functionality
1 – Enable Baseline Clamp
0 – Disable Baseline Clamp
Return
unsigned int
DRV_SUCCESS
DRV_NOT_INITIALIZED
DRV_ACQUIRING
DRV_NOT_SUPPORTED
DRV_P1INVALID
Parameters set.
System not initialized.
Acquisition in progress.
Baseline Clamp not supported on this camera
State parameter was not zero or one.
SetBaselineOffset
unsigned int WINAPI SetBaselineOffset(int offset)
Description
This function allows the user to move the baseline level by the amount selected. For
example “+100” will add approximately 100 counts to the default baseline value. The
value entered should be a multiple of 100 between -1000 and +1000 inclusively.
Parameters
Int offset: Amount to offset baseline by
Return
unsigned int
DRV_SUCCESS
DRV_NOT_INITIALIZED
DRV_NOT_AVAILABLE
DRV_ACQUIRING
DRV_P1INVALID
Parameters set
System not initialized
Baseline Clamp not available for this camera
Acquisition in progress
Offset out of range
NOTE
Only available on iXon range