User's Manual

NaviCam Small Bowel Capsule Endoscopy System
User Manual P a g e |
74
ANKON Technologies Co., Ltd. AKQWO-YF-018 Rev A/1
Figure 6-3 AKR-1 Data Recorder Operation Manual
①: Alarm indicator, yellow; ⑦: Powerswitch;
②: Run indicator, green; ⑧: USB interface to connect
③④⑤: Battery capacity indicators; ⑨: Charging port.
⑥: Charge indicator,white;
Hold down the power switch without releasing until all the five LEDs (marked as
①②③④⑤ in Figure 6-3) are lit up to power on the Data Recorder normally. Click the related
button in the ESView to power the Data Recorder off. If the Data Recorder is not connected to the
computer via USB, and therefore no capsule information is detected for a consecutive 15 minutes,
it will automatically power off to save energy. If the yellow LED alarm indicator (marked as in
Figure 6-3 with word “alarm under it) is always on, it indicates that no capsule is detected. If
on/off continues flashing, it indicates that the capsule is present, however the upper-level
software has not yet issued a command to let the capsule capture images. The green LED run
indicator (marked as in Figure 6-3 with word “run” under it) will be lit up once it receives the
image information from the capsule and off again quickly; it will light up again once it receives
the next image information and then off again. The received image information captured by the
capsule will also be marked accordingly.
The green LED battery capacity indicators (marked as ③④⑤ in Figure 6-3, with battery
capacity icon under it, three lattices, two lattices and one lattice specifically) indicate
current battery capacity. It is 90% when ③④⑤LEDs are all lit up, more than 70% when
④⑤ LEDs are lit up, and 40% when LED is lit up. If only LED is on/off flashing, the
remaining battery capacity is very limited. If the Data Recorder is powered on and being charged,