To power off the AKRI-1 Data Recorder, click the Turn On/Off button in ESView Software
Interface.
If the AKRI-1 Data Recorder cannot detect the information of the capsule for 30 minutes and
cannot connect to the software of Rtdisplay via USB port, it would be automatically power
off to save power.
If the telecommunication indicator light flashes green, it means the capsule’s data are
adequately received.
If the telecommunication indicator light flashes yellow, it means some abnormalities
occurred, but the capsule’s data has been received.
If the telecommunication indicator light does not flash, it means no capsule’s data have been
received.
The Status Indicator Light is should be green. It will turn yellow in the following situations:
(1) Low Battery; (2) Belt not Connected; (3) Other Equipment Exception.
The Charging Indicator Light remains on when power is on and turns off if power is
disconnected. If the Data Recorder is not full charged, the Charging Indicator Light will flash
yellow, if it is fully charged, the Charging Indicator Light will flash green.
Press the Screen Display button to light the screen for 30s. If it is pressed again during the
display, it will count the time again to display for 30s. The information displayed on the
screen is as follows:
Row 1: (1) Telecommunication Status Icon. (2) Battery Icon: Battery charge level.
Row 2: (1) Wireless Channel. (2) The Pattern State of Small Intestine.
Row 3: Check-in state/Time of Examination. It will display “No Entry” if not entered.
Otherwise, it will display the time of examination.
Row 4: Error Information. If there have no error information, no display will appear.
The base can charge and transmit data. The Power-on Indicator Light will flash green when
the power port of the base is powered-on. It can transmit data when the base is connected to
the computer via data port.
WARNING
The Data Recorder should be charged only with provided charger.
Charging can only be performed by the physician/operator.
Patients shall not have access to the adaptor.