User's Manual

NaviCam Small Bowel Capsule Endoscopy System
User Manual P a g e |
79
ANKON Technologies Co., Ltd. AKQWO-YF-018 Rev A/1
Figure 7-2: Schematic Diagram of turning-on the Capsule
7.2.2 Locate the Capsule
To detect the location of the capsule in a patient using the Locator, perform the following steps:
1. Turn on: Hold the Locator upright, start the Locator by pressing “Power Switch in the open
space without metal objects nearby. Once the Locator has started, the working indicator will
flash at a frequency of once per second, and the scanning indicator will be off.
2. Locating: Use the Locator to move gently in the detection area. Once the Locator is near the
capsule, the scanning indicator will light up. When the scanning indicator stays on, it
indicates that the capsule is right below the scanning indicator.
7.2.3 Turn-off the Locator
Press and hold the “Power Switch for more than 5 seconds and the Locator will shut down. All the
indicators will be off after the shutdown. The Locator will automatically shut down after 10
minutes without any operation.
7.2.4 Charge the Locator
The Locator uses a battery to supply power. When the battery voltage is low, the flashing
frequency of the working indicator will get lower. When the flashing changes to one time per three
seconds, the Locator needs to charge through the power port. During the charging process, the
charging indicator is always on, and is automatically turned off after being fully charged.
WARNING
The Locator can only be charged using a charger that comes with
the unit.
Charging can only be performed by the physician/operator.
Patients shall not have access to the adaptor.