NaviCam Small Bowel Capsule Endoscopy System
User Manual P a g e |
80
ANKON Technologies Co., Ltd. AKQWO-YF-018 Rev A/1
C
HAPTER
8: TROUBLESHOOTING
8.1 ESView Warnings
Warning Message Cause Action
Fails to add patient
information
Recording fails Add patient information again
Complete addition of
patient information
Record added
Click “Confirm” to exit the dialog box
Save change of patient
information
Record changed
Click “Confirm” to exit the dialog box
Cancel change of patient
information
Confirm cancel
Click “Confirm” to confirm cancel,
or click “Cancel” to undo it
Delete patient information Confirm deletion
Click “Confirm” to confirm
deletion, or click Cancel” to
undo it
Add case completed Record added
Click “Confirm” to exit the dialog box
Delete case information Confirm deletion
Click “Confirm” to confirm deletion,
or click “Cancel” to undo it
Connection error
Cannot connect to the data
recorder, need to check if
USB is connected properly
Click “Confirm” and reconnect the
data recorder to USB
Data export warning
Data export is not
completed
Click “Yes to stop data export or “No”
to continue export
Exit video view Exit video view
Click “Yes to exit the image browser,
click “No” or “Cancel” to undo it
Confirm exit Confirm exit
Click “Confirm” to exit the software or
“Cancel” to undoit
Print error Printer error
Check printer power and printer
settings
8.2 Capsule
Problem
Cause
Action
When activating capsule,
LED light does not lit up
Battery has no power or
capsule
damaged
Replace the capsule
Warning:
replacement
of
Capsule is not activated Activate the capsule again
Equipment failure
Contact technical
support
personnel