User's Manual

NaviCam Small Bowel Capsule Endoscopy System
User Manual P a g e |
81
ANKON Technologies Co., Ltd. AKQWO-YF-018 Rev A/1
Problem
Cause
Action
Real-time image is not
transmitted to computer
Transmission problems with
Data
Recorder
See below actions
Capsule is not activated
Activate the capsule
using
the
Locator
8.3 Data Recorder
Exception
Possible
Cause
Action
Turn the power switch on,
and LED or screen are not lit
up
Low battery Charge the system for 8 hours
Device malfunction
Contact a technical support
staff
An image is not transmitted
to your computer in
real-time
The capsule is not activated
Use Locator to activate the
capsule
Signal transceiverof the
recorder is invalid
Contact a technical support
staff
8.4 Locator
Problem Possible Cause Action
Turn the power switch on,
and LED not lit up
The battery is dead
Charge the system again for 8
hours
Device malfunction
Contact technical support
staff
Cannot activate the capsule
Capsule malfunction
Contact technical support
staff
Device malfunction
Contact technical support
staff
WARNING
If problem cannot be addressed with the above actions, please
contact ANKON Technologies Customer Support.