User's Manual

NaviCam Small Bowel Capsule Endoscopy System
User Manual P a g e |
82
ANKON Technologies Co., Ltd. AKQWO-YF-018 Rev A/1
C
HAPTER
9: MAINTENANCE
9.1 Cleaning the Data Recorder Vest/Belt-Examination Cloth
After each procedure, the examination vest/belt should be sprayed with 70% IPA and then wiped
with a clean, dry cloth and air dried for at least 15 minutes. The outer layer of fabric can be
washed in water. For the vest, the inside layer that contains the antenna and cables is not
washable, it is not in direct contact with the human body and does not require cleaning.
The Data Recorder vest/belt should be cleaned as follows:
1. Take the Data Recorder out of the vest/belt.
2. Remove the outer layer of fabric and immerse it in warm water with neutral soap for 20
minutes.
3. Gently rub and then dry it.
CAUTION
Do not dry clean or iron the vest.
9.2 Cleaning the Data Recorder and Locator
The Data Recorder and Locator should be manually cleaned after every use by wiping the
equipment surface gently with a cotton ball lightly dipped in 70% isopropyl alcohol and let it dry
for 15 minutes.
Note:
As the equipment is equipped with batteries inside, please use a cotton ball dipped with 70%
isopropyl alcohol and then squeeze excessive alcohol from the cotton ball before using it for
wiping to avoid alcohol flowing into the equipment through the seams when wiping the edges of
the equipment.