Guidance and manufacturer’s declaration - electromagneticimmunity
The Data Recorder is intended for use in the electromagnetic environment specifiedbelow.
The user of the Data Recorder should assure that it is used in such anenvironment.
Immunity test IEC 60601 test
level
Compliance level Electromagnetic
environment -
guidance
Portable and mobile RF
communications equipment
should be used no closer to any
part of the Data Recorder,
including cables, than the
recommended separation
distance calculated from the
equation applicable to the
frequency of the transmitter.
Recommended separation
distance
Conducte
d RF
IEC
61000-4-
6
3 V
150 kHz 80 MHz
6 V in ISM bands
between 150
kHz
and 80 MHz
80% AM at 1 kHz
3 V
150 kHz 80 MHz
6 V in ISM bands
between 150
kHz
and 80 MHz
80% AM at 1 kHz
Mains power quality should
be that of a typical
commercial or hospital
environment.
Radiated RF
IEC
61000-4-
3
3 V/m
80 MHz 2.7 GHz
80% AM at 1 kHz
3 V/m
80 MHz 2.7 GHz
80% AM at 1 kHz
Mains power quality should
be that of a typical
commercial or hospital
environment.