Guidance and manufacturer’s declaration - electromagneticimmunity
The Capsules are intended for use in the electromagneticenvironment specified below.
The user of the Capsules should assure that it is used insuch an environment.
Electrical
fast transient/burst
IEC 61000-4-4
±2 kV for power
supply Lines
±1 kV for input/output
Lines
N/A N/A
Surge IEC/EN
61000-4-5
±0.5 kV, ±1 kV line(s)
to line(s)
±0.5 kV, ±1 kV, ±2 kV
line(s) to earth
N/A N/A
Voltage dips, short
interruptions
and voltage
Voltage dips
0% U
T
; 0.5
cycle
N/A N/A
variations on
power supply
input Lines
IEC 61000-4-11
At 0°, 45°, 90°,
135°,180°, 225°, 270°
and 315°
0% U
T
; 1 cycle
and
70% U
T
; 25/30
cycles
Single phase:
at
Voltage
interruptions:
0% U
T
; 250/300
cycle
Power frequency
(50/60 Hz) magnetic
field
IEC 61000-4-8
30 A/m
50 Hz or 60 Hz
30 A/m
50 Hz or 60 Hz
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical commercial
or hospital environment.
NOTE: U
T
is the AC mains voltage prior to application of the test level.