User's Manual

NaviCam Small Bowel Capsule Endoscopy System
User Manual P a g e |
93
ANKON Technologies Co., Ltd. AKQWO-YF-018 Rev A/1
Guidance and manufacturer’s declaration - electromagneticimmunity
The Capsules are intended for use in the electromagneticenvironment specified below.
The user of the Capsules should assure that it is used insuch an environment.
Immunity test IEC 60601 test
level
Compliance level Electromagnetic
environment -
guidance
Portable and mobile RF
communications equipment
should be used no closer to any
part of a Capsule, including
cables, than the recommended
separation distance calculated
from the equation applicable
to the frequency of the
transmitter.
Recommended separation
distance
Conducted RF
IEC 61000-4-6
3 V
150 kHz 80 MHz
6 V in ISM bands
between 150 kHz
and 80 MHz
80% AM at 1 kHz
N/A N/A
Radiated RF
IEC
61000-4-
3
3 V/m
80 MHz 2.7 GHz
80% AM at 1 kHz
3 V/m
80 MHz 2.7 GHz
80% AM at 1 kHz
Mains power quality should
be that of a typical
commercial or hospital
environment.