Guidance and manufacturer’s declaration - electromagneticimmunity
The Locator is intended for use in the electromagnetic environment specified below.
The user of the Locator should assure that it is used in such anenvironment.
Surge
IEC/EN
61000-4-5
±0.5 kV, ±1 kV line(s) to
line(s)
±0.5 kV, ±1 kV, ±2 kV
line(s) to earth
±0.5 kV, ±1 kV, ±2 kV
line(s) to earth
Mains power quality
should be that of a
typical commercial or
hospital environment.
Voltage dips,
short
interruptions and
voltage variations
on power supply
inputLines
IEC 61000-4-11
Voltage dips
0% U
T
; 0.5
cycle
At 0°, 45°, 90°, 135°,
180°, 225°, 270° and
315°
0% U
T
; 1 cycle
and
70% U
T
; 25/30
cycles Single phase: at
Voltage interruptions
0% U
T
; 250/300
cycle
Voltage dips
0% U
T
; 0.5
cycle
At 0°, 45°, 90°, 135°,
180°, 225°, 270° and
315°
0% U
T
; 1 cycle
and
70% U
T
; 25/30
cycles Single phase: at
Voltage interruptions
0% U
T
; 250/300
cycle
Mains power quality
should be that of a
typical commercial or
hospital environment.
If the user of the Data
Recorder requires
continued operation
during power mains
interruptions, it is
recommended that
the Locator be
powered from an un-
interruptible power
supply or a battery.
Power frequency
(50/60 Hz)
magnetic
Field
IEC 61000-4-8
30 A/m
50 Hz or 60 Hz
30 A/m
50 Hz or 60 Hz
Power frequency
magnetic fields
Should be at levels
characteristic of a
typical location in a
typical commercial or
hospital environment.
NOTE: U
T
is the AC mains voltage prior to application of the test level.