User's Manual

NaviCam Small Bowel Capsule Endoscopy System
User Manual P a g e |
96
ANKON Technologies Co., Ltd. AKQWO-YF-018 Rev A/1
Guidance and manufacturer’s declaration - electromagneticimmunity
The Locator is intended for use in the electromagnetic environment specified below.
Theof the Locator should assure that it is used in such an environment.
Immunity test
IEC 60601 test
level
Compliance level
Electromagnetic
environment - guidance
Portable and mobile RF
communications equipment
should be used no closer to
any part of the Locator,
including cables, than the
recommended separation
distance calculated from the
equation applicable to the
frequency of the transmitter.
Recommended separation
distance
Conducted
RF
IEC
61000-4-6
3 V
150 kHz 80 MHz
6 V in ISM
bands
between 150
kHz
and 80 MHz
80% AM at 1
kHz
3 V
150 kHz 80 MHz
6 V in ISM
bands between
150 kHz
and 80 MHz
80% AM at 1 kHz
Mains power quality should
be that of a typical
commercial or hospital
environment.
Radiated RF
IEC
61000-4-
3
3 V/m
80 MHz 2.7 GHz
80% AM at 1 kHz
3 V/m
80 MHz 2.7 GHz
80% AM at 1 kHz
Mains power quality should
be that of a typical
commercial or hospital
environment.