Step 4: Now click on the measurement icon at the top of the page and select “Calibration.”
Step 5: The software will then prompt you to take a measurement on the selected calibration
standard.
Step 6: Take a sample measurement of the selected calibration standard by clicking on
beginning and ending points across a known measurement value (such as 5 mm, .5 in., etc.). A
measurement value before calibration will be displayed (5.179 mm in the example below).