User manual

30
Step 7: Enter the exact known value of the measurement sample (5.0 mm in this example) in
the “Adjusted Mag” window. Click “OK.”
Step 8: The software will automatically calibrate the magnification rate and will show the
adjusted magnification rate in the upper right corner of the picture.
Step 9: Now the software is ready to take calibrated measurements with adjusted magnification
rates. To verify calibration simply take another measurement sample of a known measurement
value.
Notes:
1. Calibration can be done at both lower magnification rates (<80x) and higher magnification
rates (>80x) respectively. However, it will be more accurate if the calibration is done at the
magnification rate closest to the one used for measurement.
2. The calibration can be cleared by selecting “Clear All Calibration.”