Specifications

NETWORK ANALYZERS
366 For product ordering information, see pages 4 – 10
Frequency range (GHz)
Test port characteristics
Specifications
System performance
Waveguide designation
Q-Band
(WR-22)
33 to 50
+10
93
.02
+7
100
V-Band
(WR-15)
50 to 75
+8
90
.05
+7
97
E-Band
(WR-12)
60 to 90
+8
90
.06
+6
96
Extended
E-Band
56 to 60
60 to 85
85 to 94
+8
85
90
76
.08
+5
+6
+4
90
96
80
W-Band
(WR-10)
75 to 100
100 to 110
+6
90
90
.06
+5
+2
95
92
Extended
W-Band
65 to 75
75 to 100
100 to 110
+6
90
89
87
.08
5
+5
+2
85
94
89
Maximum signal into
port 2 (dBm)
Noise floor (dBm)
103 98 98
93
98
84
96
96
96
95
93
Receiver dynamic
range (dB)
1
High level noise
(dB, typical)
Power @ DUT
(dBm, typical)
System dynamic
range (dB)
2
1: Receiver dynamic range is defined as the ratio of the maximum signal level at Port 2 for 0.1 dB compression to the system noise floor.
2: System dynamic range is defined as the ratio of the power at Port 1 and the system noise floor (forward measurements only).
Offset short calibration
1
Waveguide designation
Q-Band
(WR-22)
V-Band
(WR-15)
E-Band
(WR-12)
Extended
E-Band
W-Band
(WR-10)
Extended
W-Band
Frequency (GHz) 33 to 50
>50
>45
>50
±0.010
±0.010
>100
50 to 75
>50
>37
>50
±0.030
±0.060
>90
60 to 90
>46
>36
>46
±0.040
±0.060
>90
56 to 94
>44
>33
>44
±0.080
±0.1
>80
75 to 110
>46
>36
>46
±0.040
±0.070
>90
65 to 110
>40
>30
>40
±0.080
±0.1
>80
Directivity (dB)
Source match (dB)
Load match (dB)
Reflection frequency tracking (dB)
Transmission frequency tracking (dB)
Isolation (dB)
1: At 23 ±3˚C using the offset short calibration method with a sliding load or LRL calibration method (as noted) to achieve 12-term error correction.
LRL calibration
1
Waveguide designation
Q-Band
(WR-22)
V-Band
(WR-15)
E-Band
(WR-12)
Extended
E-Band
W-Band
(WR-10)
Extended
W-Band
Frequency (GHz) 33 to 50
>50
>50
>50
±0.002
±0.002
>100
50 to 75
>50
>50
>50
±0.002
±0.002
>90
60 to 90
>46
>46
>46
±0.002
±0.002
>90
56 to 94
>44
>43
>44
±0.006
±0.006
>80
75 to 110
>46
>46
>46
±0.002
±0.002
>90
65 to 110
>40
>40
>40
±0.006
±0.006
>80
Directivity (dB)
Source match (dB)
Load match (dB)
Reflection frequency tracking (dB)
Transmission frequency tracking (dB)
Isolation (dB)
F-Band
(WR-8)
90 to 115
115 to 140
+4
-88
-87
.08
3
7
85
80
92
91
F-Band
(WR-8)
90 to 140
>45
>34
>45
±0.060
±0.1
>80
F-Band
(WR-8)
90 to 140
>45
>45
>45
±0.004
±0.004
>80