Specifications

http://www.anritsu.com 383
NETWORK ANALYZERS
7
Transmission characteristics (ratio measurement): Amplitude, phase, group delay
Measurement items Reflection/impedance characteristics: Amplitude, phase (with external transducer)
Level characteristics: Absolute amplitude
Range: 10 Hz to 300 MHz
Resolution: 0.01 Hz
Accuracy (standard)
Frequency Aging rate: 1 x 10
–6
/day (15 minutes after power-on)
Temperature characteristics: ±5 x 10
–6
(0˚ to 50˚C)
Accuracy (Option 13: High-stability reference oscillator)
Aging rate: ±2 x 10
–8
/day (24 h after power-on)
Temperature characteristics: ±5 x 10
–8
(0˚ to 50˚C)
Channel No.
Standard: 2 (R, TA); Option 12: 3 (R, TA, TB)
Impedance: 50 , 1 M switchable (when combined with MA4605A: 75 , 1 M)
Input range (IRG): 0/+20 dBm
Input Max. input power
AC: +20 dBm; DC ±2.2 V (50 )
AC: 0 dBm; DC: ±20 V (1 M)
Connector: BNC-J
Probe source: +12 ±1 V, 100 mA (with protective circuit for shorts)
Average noise level –120 dBm (RBW: 1 kHz, 1 to 300 MHz), –110 dBm (RBW: 1 kHz, 80 kHz to 1 MHz)
Crosstalk
Between channels: 120 dB (80 kHz to 300 MHz), 110 dB (up to 80 kHz)
Between transmitter and receiver: 125 dB
Resolution bandwidth 3, 10, 30, 100, 500 Hz, 1, 2, 3, 4, 5, 10, 20 kHz and automatic setting
Output level range
Output A: 0 to +21 dBm; Option 10: –70 to +21 dBm
Output B: –6 to +15 dBm (–9.5 to +11.5 dB when Option 14 added);
Option 10: –76 to +15 dBm (–79.5 to +11.5 dB when Option 14 added)
Output resolution: 0.01 dB
Output Output level accuracy: ±1.0 dB (frequency: 100 MHz, Output A: +10 dBm)
Output level linearity: ±0.5 dB (0 dBm reference, frequency: 100 MHz, Output A: 0 to +21 dBm)
Output level deviation: ±1.5 dB (output A: +10 dBm, 100 MHz reference)
Step error: ±0.5 dB (Option 10)
Output impedance: 50 (when combined with MA4605A: 75 )
Connector: BNC-J
Measurement range: 120 dB
Measurement resolution: 0.001 dB
Display scale: 0.01 dB/div to 50 dB/div (1-2-5 sequence)
Dynamic accuracy
Amplitude
measurement
Level relative to IRG 80 kHz to 100 MHz 10 kHz to 300 MHz
0 to –10 dB ±0.30 dB ±0.30 dB
–10 to –60 dB ±0.05 dB ±0.05 dB
–60 to –70 dB ±0.10 dB ±0.30 dB
–70 to –80 dB ±0.30 dB ±1.00 dB
–80 to –90 dB ±1.20 dB ±4.00 dB
–90 to –100 dB ±4.00 dB
Specifications
Resonator 2 measurement
Continued on next page