Specifications
NETWORK ANALYZERS
352 For product ordering information, see pages 4 – 10
Specifications
Standard connector type N female
Optional connector types 3.5 mm female, 3.5 mm male, GPC-7, N male
Connector
Frequency Directivity Source match Load match
(MHz) (dB) (dB) (dB)
Measurement port
characteristics
Frequency range
MS4622A/B, 10 MHz to 3 GHz
MS4623A/B, 10 MHz to 6 GHz
Frequency resolution 1Hz
Frequency stability (with
internal time base) – aging
<5x10
-6
/ year
Temperature <5x10
-6
over +15°C to +50°C
MS4622A Transmission/Reflection Test Set +10 to –85 dBm
MS4622B Active Reversing Test Set +10 to –85 dBm
MS4622B (Opt 3) w/ 2nd Source, 3rd Test Port & S/A. +10 to –85 dBm
MS4622B (Opt 4) w/ Noise Figure +7 to –85 dBm
Power output range MS4622B (Opt 6) w/ 3rd Test Port +10 to –85 dBm
MS4623A Transmission/Reflection Test Set +10 to –85 dBm
MS4623B Active Reversing Test Set +7 to –85 dBm
MS4623B (Opt 3) w/ 2nd Source, 3rd Test Port & S/A +7 to –85 dBm
MS4623B (Opt 4) w/ Noise Figure (3 GHz only) +5 to –85 dBm
MS4623B (Opt 6) w/ 3rd Test Port +7 to –85 dBm
Power control range ≥ 20 dB. The minimum absolute level for power sweep is –15 dBm while the maximum power output for a unit is +10 dBm.
The source power (dBm) may be set from the front panel menu or via GPIB. Port 1 power level is settable from
Source power level +10 dBm (on the simpler test sets, ranging to +5 dBm on the most complex) to –15 dBm with 0.01 dB resolution.
In addition, the port 1 (& port 3) power may be attenuated in 10 dB steps using the internal 70 dB step attenuator.
Power level accuracy ±1 dB (no flat power calibration applied; full-band frequency sweep at –15 dBm, 0 dBm, and maximum rated power)
Level test port power The power at all sweep frequencies is leveled to within ±1dB. Only port 1 and port 3 (if installed) can be externally leveled.
Harmonics and spurious <–30 dBc at maximum rated power
Sweep type Linear, CW, Marker, or N-Discrete point sweep
Power sweep range 20 dB (minimum)
Frequency range 10 MHz to 3 GHz (6 GHz)
Frequency
resolution
1 Hz
Source
Power level
#2
accuracy
±1 dB (no flat power calibration applied; full-band frequency sweep at –15 dBm, 0 dBm & max rated power)
(optional) Harmonics
and spurious
<–30 dBc at maximum rated power (not applicable in harmonic measurement mode)
Sweep type Linear, CW, Marker, or N-Discrete point sweep
Power sweep
range
20 dB (minimum)
Test port characteristics
10 to 1000
∗1,2
1000 to 3000
∗1,2
3000 to 6000
∗1,2
10 to 1000
∗1,3
1000 to 3000
∗1,3
3000 to 6000
∗1,3
10 to 1000
∗4
1000 to 3000
∗4
3000 to 6000
∗4
>46
>44
>38
>44
>42
>37
>30
>25
>20
>44
>41
>39
>42
>40
>37
>15
>15
>15
>46
>44
>38
>44
>42
>37
>15
>15
>15
3.5 mm
10 to 1000
∗1,2
1000 to 3000
∗1,2
3000 to 6000
∗1,2
10 to 1000
∗1,3
1000 to 3000
∗1,3
3000 to 6000
∗1,3
10 to 1000
∗4
1000 to 3000
∗4
3000 to 6000
∗4
>46
>44
>38
>44
>42
>37
>30
>25
>20
>44
>41
>39
>42
>40
>37
>15
>15
>15
>46
>44
>38
>44
>42
>37
>15
>15
>15
N-Type
10 to 1000
∗1,2
1000 to 3000
∗1,2
3000 to 6000
∗1,2
10 to 1000
∗1,3
1000 to 3000
∗1,3
3000 to 6000
∗1,3
10 to 1000
∗4
1000 to 3000
∗4
3000 to 6000
∗4
>46
>44
>38
>44
>42
>37
>30
>25
>20
>44
>41
>39
>42
>40
>37
>15
>15
>15
>46
>44
>38
>44
>42
>37
>15
>15
>15
GPC-7
Source specifications
∗1: 12-term error correction applied
∗2: Port 1 & Port 2
∗3: Optional Port 3 added
∗4: Uncorrected performance
Continued on next page