Specifications

Menu Name Function Description GPIB Mnemonic
SOURCE 1
SETUP
SOURCE 1 POWER
XX.XX dBm
(5 TO -15 dB)
Enter and display the power level, in dBm, of the
Source 1 Power Level.
PWR; PWR?
PORT1 ATTN:
0*10 dB (0-70 dB)
Attenuates the microwave source power at Port 1
from 0 to 70 dB, in 10 dB steps. The power is
attenuated before being applied to Port 1 for a
forward transmission or reflection test (S21 or S11,
respectively).
SA1; SA1?
PORT1 POWER
XX.XX dBm
Displays the net port power: Sum of Source 1 power
and attenuator settings.
P1P?
FLAT TEST PORT
POWER CAL
((NO) CAL EXISTS)
Changes the soft-key display to the FLAT TEST
PORT POWER CAL menu.
(FP1DONE?)
FLATNESS
CORRECTION
ON/OFF
AT XX.XX dBm
Toggles power flatness correction on or off and
displays the correction level.
FP1; FP0; FPX?
RETURN Returns to the previous soft-key menu. None
SOURCE 2 FLAT
TEST PORT
POWER
CALIBRATION
XXX POINTS
MEASURE 1 PWR
POINT EVERY XX
POINT(S)
Number of points between measurement of power
taken during flat test port power calibration.
PTS3; PTS3?
POWER TARGET
XX.XX dBm
Power level at which calibration will try to level. PTP3; PTP3?
BEGIN CAL Starts the calibration. TP3; SFC; TPX?
RETURN Returns to the previous soft-key menu. None
SOURCE 2
SETUP
EXTERNAL
SOURCE
SOURCE 2 POWER
XX.XX dBm
Sets external Source 2 power level. PW2; PW2?
RETURN Returns to the previous soft-key menu. None
SOURCE 2
SETUP INTERNAL
SOURCE
SOURCE 2 POWER
XX.XX dBm
(5 TO -15 dB)
Sets Source 2 power level. PW2; PW2?
PORT 3 ATTN
0*10 dB
(0 TO -70 dB)
Sets Port 3 attenuator in 10 dB increments. SA3; SA3?
PORT3 POWER
XX.XX dBm
Displays the net port power: Sum of Source 2 power
and attenuator settings.
P3P?
FLAT TEST PORT
POWER CAL
((NO) CAL EXISTS)
Changes the soft-key display to the SOURCE 2
FLAT TEST PORT POWER CALIBRATION menu.
None
FP3DONE?
FLATNESS
CORRECTION
ON/OFF
AT XX.XX dBm
Toggles power flatness correction on or off. FP30; FP31; FP3X?
RETURN Returns to the previous soft-key menu. None
SOURCE 3
SETUP
SOURCE 3 POWER
XX.XX dBm
Sets external Source 3 power level. PW3; PW3?
RETURN Returns to the previous soft-key menu. None
FRONT PANEL POWER
OPERATION KEY MENUS
MS462XX OM 4-75